For companies developing and manufacturing medical devices for international markets, the MDSAP represents an important reference for auditing their quality management system.
On August 3, 2026, the MDSAP Audit Approach Rev. 11, identified as MDSAP AU P0002.011, was published. The new revision updates the methods used by auditors to verify various aspects of the Quality Management System, with particular attention to certain issues that directly involve R&D, Technical Management, Product Development, and Production.
It is important to clarify one point immediately: Rev. 11 does not introduce a new standard and does not modify ISO 13485. Furthermore, the MDSAP Audit Approach does not add regulatory requirements for manufacturers. The document is intended for auditing organizations to systematically verify requirements already present in ISO 13485 and the applicable regulations of the countries participating in the program.
The useful question for a manufacturer therefore becomes: which aspects of the QMS and the development and production processes deserve greater attention today in view of an MDSAP audit?
What is MDSAP and why does it also concern R&D and Technical Management?
The Medical Device Single Audit Program (MDSAP) allows an organization’s quality system to be subjected to a single audit that takes into account the requirements of participating regulatory authorities, including Australia, Brazil, Canada, Japan, and the United States.
The audit follows a process-based approach. The four main processes concern Management, Measurement Analysis and Improvement, Design and Development, and Production and Service Controls. The Purchasing process is linked to these, while Risk Management forms a cross-cutting foundation for the entire system.
This aspect is particularly important.
An MDSAP audit not only verifies the existence of procedures and records. The auditor also examines the connections between processes: a decision made during design can impact risk management, production, controls, suppliers, and, subsequently, information gathered from the market.
As a result, the MDSAP does not exclusively concern Quality and Regulatory Affairs. A significant portion of the evidence comes from the daily activities of R&D, design, industrialization, and production.
MDSAP Rev. 11: What are the most interesting changes?
Rev. 11 contains numerous updates and clarifications. However, not all of these are equally relevant to those involved in product development.
For our field, four aspects deserve particular attention: the risk-based approach in the QMS, personnel skills, integration between design and risk management, and cybersecurity where applicable.
Furthermore, the revision updates some specific requirements related to different markets and further clarifies aspects related to product, production, and supplier changes.
Let’s see what operational implications these changes may have.
MDSAP and risk management: risk must truly enter into processes
One of the most significant updates concerns the Management process.

Rev. 11 moves the verification of the application of a risk-based approach to QMS process control to Management Task 1. At the same time, Task 7 focuses more on overseeing risk management throughout the product life cycle.
The distinction is important.
On the one hand, we find the risk management of the device, including hazard identification, risk assessment, and related control measures. On the other, we find the use of risk to define and control quality system processes.
In practice, simply having a formally correct Risk Management File is not enough. The organization must be able to demonstrate that the information from the risk analysis actually influences design, production, controls, suppliers, and change management.
What impact on product development?
Consider replacing a material, modifying a functional geometry, or changing a manufacturing process.
The development team should not evaluate these changes solely from a technical or economic perspective. They must also assess whether they introduce new risks, modify already identified risks, or affect a control measure.
Therefore, risk management and product development must proceed together.
This principle becomes particularly important during industrialization, when design choices begin to transform into definitive materials, processes, tolerances, manufacturing parameters, and control criteria.
Skills: it is no longer enough to demonstrate that a person has attended a course
Another interesting change concerns Management Task 6.
Rev. 11 shifts the focus from simple training to personnel competence. The official summary of the changes defines this intervention as a shift in focus from “training” to “competence.”
The difference is tangible.
A registration proves that a person has attended a course. However, it does not automatically prove that that person possesses the skills required to correctly perform a given task.
For this reason, it is important to establish a clear relationship between:
role → required skills → training or experience → verification of competence.
This principle applies to designers and development managers, but also to production operators, quality control personnel, validation personnel, and process managers who can influence device quality.
For companies introducing new technologies or new production processes, this aspect deserves particular attention.
MDSAP, design and industrialization: connected evidence is needed
The Design and Development process remains a central focus of the Audit Approach.
The MDSAP verifies the relationship between development planning, design inputs, design outputs, verification, validation, risk management, change management, and design transfer.
Therefore, it’s not just the individual documents that matter.
An initial requirement must be addressed by the design outputs. Verification activities must demonstrate that the outputs meet the inputs. Validation must confirm, when applicable, that the device meets the user’s needs and intended use. At the same time, risk management must inform these decisions.
From design to production
The connection becomes even more evident during design transfer.
If, during development, a dimensional characteristic or a specific material is critical to safety or performance, this information must be correctly transmitted to manufacturing.
The same applies to any process parameters, controls, test methods, and acceptance criteria.
Therefore, industrializing a medical device does not simply mean “putting an approved CAD model into production.” It means transforming development outputs into a process capable of producing compliant and repeatable results.

Involving industrialization skills early in the development phase can therefore reduce late changes, rework, and difficulties during verification and validation.
MDSAP and cybersecurity: when development and post-market must communicate
For devices incorporating software, firmware, or programmable logic, Rev. 11 significantly strengthens the cybersecurity focus.
The revision adds specific FDA references to the Design and Development process and introduces cybersecurity guidance to Measurement, Analysis, and Improvement activities, linking them to information collected in the post-market phase.
The principle is clear: cybersecurity does not end with the initial design.
New vulnerabilities, incidents, or information collected after marketing can alter the risk landscape. This information must therefore be incorporated into risk management processes and, when necessary, lead to product modifications.
For R&D, this means better integrating mechanical, electronic, software, quality, and regulatory expertise. Naturally, the importance of this topic varies greatly depending on the device. For a purely mechanical product, it may be of little relevance, while it becomes crucial for devices with software or connectivity.
Suppliers and outsourced processes remain under the control of the manufacturer
Supplier relationships are another important element of the MDSAP approach.
When an organization outsources a process or purchases a product that could affect device compliance, it must maintain adequate controls over the supplier and the outsourced process. Responsibility does not disappear simply because an activity is entrusted to an external partner.
This principle directly affects companies that outsource activities such as design, prototyping, tooling, molding, machining, assembly, or specialized testing.

Therefore, clarity of specifications, acceptance criteria, responsibilities, change management, and records demonstrating what has actually been achieved become particularly important.
The same underlying theme emerges here: the QMS must be connected to actual technical activities.
What should a company interested in MDSAP check today?
The publication of Rev. 11 does not mean that a manufacturer must automatically rewrite its quality system.
Rather, it represents an opportunity to conduct a targeted gap analysis and verify whether the available evidence clearly demonstrates the functioning of the processes. The MDSAP program itself specifies that the Audit Approach reflects existing requirements of ISO 13485 and the regulations of participating authorities, without introducing additional requirements.
Before the next audit, it may therefore be useful to verify:
whether the risk-based approach is effectively applied to the QMS processes;
whether risk management and design remain linked throughout development;
whether the organization demonstrates personnel competence beyond simple training;
whether design input, output, verification, validation, and risk management are consistent and traceable;
whether technical changes are also assessed for risk;
whether design transfer transfers all necessary information to manufacturing;
whether supplier controls are proportionate to their influence on device compliance;
Whether, when applicable, cybersecurity and post-market information enter the development and improvement process.
The final question should therefore not be “how many documents do we need to add?”, but “how well can we demonstrate the connection between what we’ve designed, evaluated, produced, and controlled?“
MDSAP Rev. 11: What’s Really Changing for Medical Device Developers?
The MDSAP Audit Approach Rev. 11 does not represent a revolution in the quality system. However, it reinforces an important principle: an effective QMS cannot exist separately from technical activities.
Risk management, product development, industrialization, production, and supplier management must communicate with each other.

For R&D and Technical Management, this means being able to reconstruct the rationale for key design decisions and demonstrate how they were verified, transferred to production, and monitored.
For Quality, it means verifying that these activities generate consistent and traceable evidence.
Ultimately, Rev. 11 confirms a direction already clear in the MDSAP approach: less focus on individual documents and more attention to the entire system’s ability to function as a set of truly connected processes.